Quickly configure, start and stop tests on the 1550C remotely, a safe distance away from any operating, ... More ...
- Quickly configure, start and stop tests on the 1550C remotely, a safe distance away from any operating, energized equipment, using Fluke Connect® and a smart phone
- Real time trending gives you easy-to-see visual cues, rather than hard to read, rapidly changing values
- Report creation using Fluke Connect desktop saves time and makes test results easier to understand
- Fluke Connect saves test results to the cloud for analysis and preventive maintenance
Table 1. Required Equipment Equipment Minimum Required Characteristics Recommended Model HV Probe 6 kV, ±1 % (1000:1 Divider) 11 kV, ±2 % for the 1555 Fluke 80K-6 80K-15 Digital Multimeter 500 mVdc to 1 V: ±0.02 % Fluke 8508 Load with Guard Terminal[1] Resistances 200 kΩ, ±1.25 %, 500 V 500 kΩ, ±1.25 %, 500 V 1 MΩ, ±1.25 %, 1 kV 2.5 MΩ, ±1.25 %, 2.5 kV 5 MΩ, ±1.25 %, 5 kV 10 MΩ, ±1.25 %, 10 kV 1 GΩ, ±1.25 %, 10 kV 100 GΩ, ±1.25 %, 10 kV 200 GΩ, ±1.25 %, 10 kV 500 GΩ, ±5 %, 10 kV 1 TΩ, ±5 %, 10 kV 2 TΩ, ±5 %, 10 kV Combinations of: Welwyn F Series, Welwyn MFP2 Series And Vishay HTS-523 Capacitors with Bleeder Resistors[2] 0.1 μF, ±5 %, 500 V, Polypropylene 1 μF, ±5 %, 2.5 kV, Polypropylene
Calibrator DC current: 2 mA Accuracy: ±1.25 % DC Voltage: 0 - 550 V Accuracy: ±0.005 % AC Voltage: 0 - 240 V, 60 Hz Accuracy: ±1.25 % Fluke 5080, Fluke 5520A IR Cable Assembly Fluke P/N 2166275 Calibration Software[3] Snorre Ammeter Fluke 8508 Personal computer IBM compatible, with Microsoft Windows XP SP2 or later + .NetFramework 2.0 or later